Gamma Spectrometry Systems with HPGe Detector

RJ46​

Gamma Spectrometry Systems with HPGe Detector

    •    Supports dual spectrum measurement of energy spectrum and time spectrum
    •    With passive efficiency calibration software
    •    Automatic pole-zero and zero dead-time correction
    •    With particle information and energy spectrum information

 

Gamma Spectrometry Systems with HPGe Detector

Product Introduction :

RJ46 Gamma Spectrometry Systems with HPGe Detector mainly includes a new type of high-purity germanium low-background spectrometer independently developed. The spectrometer adopts a particle event readout method, and uses digital multi-channel to obtain and store the energy (amplitude) and time information of the HPGe detector output signal.

 

System composition:

The RJ46 Gamma Spectrometry Systems measurement system is mainly composed of three parts: high-purity germanium detector, multi-channel signal processor, and lead chamber. The detector array includes the HPGe main detector, digital multi-channel pulse processor, and low-noise high and low voltage power supply; the host computer software mainly includes parameter configuration module, particle event information receiving module, coincidence/anti-coincidence measurement module, and spectrum line display module.

 

Features:

① Supports dual spectrum measurement of energy spectrum and time spectrum

② Data can be transferred via Ethernet and USB

③ With passive efficiency calibration software

④ High time and high energy resolution, high throughput support

⑤ Digital filter shaping, automatic baseline subtraction

⑥ Able to receive particle information and energy spectrum information transmitted by the device and save it as a database

⑦ Meets the following standards :

《Gamma Spectroscopy Analysis Method for Radionuclides in Biological Samples》 GB/T 1615-2020

《Gamma Spectroscopy Analysis Method for Radionuclides in Water》 GB/T 16140-2018

《General Method for Gamma Spectroscopy Analysis of High Purity Germanium》 GB/T 11713-2015

《γ-ray spectrum analysis method for radionuclides in soil》”GB T 11743-2013

《Gamma Spectrum Analysis Method for Radionuclides in Air》 WS/T 184-2017

《Ge Gamma-ray Spectrometer Calibration Specification》JJF 1850-2020

《Technical Specification for Gamma Nuclide Measurement of Environmental Samples in Emergency Monitoring》 HJ 1127-2020

 

Main technical indicators :

Detector :

① Crystal type: High purity germanium

② Energy response range: 40keV~10MeV

③ Relative efficiency: ≥60%

④ Energy resolution: ≤2keV for 1.332 MeV peak; ≤1000eV for 122keV peak

⑤ Peak to Compressor Ratio: ≥68:1

⑥ Peak shape parameters: FW.1M/FWHM≤2.0

Digital Multi-Channel Analyzer :

① Maximum data throughput rate: no less than 100kcps

② Gain: Set coarse and fine adjustment to meet the adjustment requirements of the spectrum amplification function

③ Compatible with charge sensitive preamplifiers, current preamplifiers, voltage preamplifiers, reset type preamplifiers, self-discharge type preamplifiers, etc.

④ Supports dual spectrum measurement of energy spectrum and time spectrum

⑤ Provides standard DB9 preamplifier power supply, compatible with NIM slot

⑥ Four transmission modes: Raw Pulse View, Shaped View, Line View, and Particle Mode

⑦ Particle mode supports measuring the arrival time, energy, rise time, fall time and other information of ray events (customizable on demand)

⑧ Supports 1 main detector signal input and up to 8 independent coincidence channel inputs

⑨ 16-bit 80MSPS, ADC sampling, can provide up to 65535 spectral lines support

⑩ High time and high energy resolution, high throughput support

⑪ Programmable high voltage and display

⑫ Data can be transferred via Ethernet and USB

⑬ Digital filter shaping, automatic baseline subtraction, ballistic loss correction, low frequency noise suppression, automatic optimization, automatic pole zero, zero dead time correction, gated baseline restoration and virtual oscilloscope function

⑭ With GammaAnt spectrum analysis and processing software, it can realize functions such as nuclide identification and sample activity measurement.

Low background lead chamber :

① The lead chamber is an original integrated casting

② Lead thickness ≥10cm

Spectrum analysis and acquisition software :

① It can acquire spectrum, set parameters, etc.

② Able to receive particle information and energy spectrum information transmitted by the device and save it as a database

③ Spectral line data processing function can realize particle and energy spectrum data analysis, processing and viewing, and support data merging, screening and splitting functions

④   With passive efficiency calibration software and probe characterization


Post time: Apr-15-2025